SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY (V6 & V8)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Baxter Healthcare Corporation
711 Park Ave.
Medina
14103
New York
UNITED STATES
SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY (V6 & V8)35700BAX

Introduction

Affected Products

Reason

Baxter Corporation is issuing a device correction for the SIGMA Spectrum Infusion Pump with Master Drug Library (Spectrum Infusion Pump) in order to replace speakers installed in pumps that have an increased potential to fail prematurely. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum infusion pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.

Before using an affected spectrum infusion pump, users should confirm the generation of an audible tone, as instructed in the device operator manual.