REPLY and ESPRIT PACEMAKERS (2016-09-30) Sep 30, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : Hospitals, General Public, Healthcare Professionals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number More than 10 numbers, contact manufacturer. F. REPLY 200 SR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY REPLY 200 SR More than 10 numbers, contact manufacturer. D. ESPRIT SR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY ESPRIT SR More than 10 numbers, contact manufacturer. E. REPLY 200 DR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY REPLY 200 DR More than 10 numbers, contact manufacturer. B. REPLY DR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY REPLY DR More than 10 numbers, contact manufacturer. C. ESPRIT DR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY ESPRIT DR More than 10 numbers, contact manufacturer. A. REPLY SR PACEMAKER Manufacturer Sorin Group Italia S.R.L. (Sorin CRM) Via Crescentino S.N. Saluggia, Vercelli 13040 ITALY REPLY SR
Affected Products
REPLY SR PACEMAKER
REPLY DR PACEMAKER
ESPRIT DR PACEMAKER
ESPRIT SR PACEMAKER
REPLY 200 DR PACEMAKER
REPLY 200 SR PACEMAKER
When an implanted device approaches the recommended replacement time (RRT), the residual longevity estimate provided by the programmer at follow-up may be overestimated, which could result in reaching RRT between two follow-up visits. Until a new programmer software version is approved, LivaNova is providing a number of patient management recommendations.