Health Products

VITROS Immunodiagnostic Products Intact PTH (Reagent Pack & Calibrators)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Ortho-Clinical Diagnostics
Felindre Meadows, Pencoed, Bridgend, Great Britain
CF35 5PZ
UNITED KINGDOM
VITROS Immunodiagnostic Products Intact PTH (Reagent Pack & Calibrators)680 2892

Introduction

Affected Products

Reason

Ortho confirmed that results obtained from VITROS Intact Parathyroid Hormone (iPTH) Reagent Packs are positively biased (i.e., falsely elevated) compared to an alternative commercially available method.

Ortho observed a positive bias of 40% for samples with iPTH concentrations <100pg/ml when testing with VITROS PTH Reagent Packs in comparison to the Roche Elecsys PTH test.

Ortho data demonstrated that this bias was consistent for all in-date lots.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime