Health Products

Various Micro Therapeutics Inc. catheters and devices (2016-10-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Micro Therapeutics Inc.
DBA EV3 Neurovascular
9775 Toledo Way
Irvine
92618
UNITED STATES
D. MARATHON FLOW DIRECTED MICRO CATHETER105-5055
A090167
Manufacturer
Micro Therapeutics Inc.
DBA EV3 Neurovascular
9775 Toledo Way
Irvine
92618
UNITED STATES
E. PIPELINE EMBOLIZATION DEVICE (25-35MM)FA-71375-25
More than 10 numbers, contact manufacturer.
Manufacturer
Micro Therapeutics Inc.
DBA EV3 Neurovascular
9775 Toledo Way
Irvine
92618
UNITED STATES
B. X-CELERATOR EXCHANGE HYDROPHILIC GUIDEWIRE - SS103-0601-300
A072386
Manufacturer
Micro Therapeutics Inc.
DBA EV3 Neurovascular
9775 Toledo Way
Irvine
92618
UNITED STATES
C. ULTRAFLOW HPC FLOW DIRECTED MICRO CATHETER105-5065
More than 10 numbers, contact manufacturer.
Manufacturer
Micro Therapeutics Inc.
DBA EV3 Neurovascular
9775 Toledo Way
Irvine
92618
UNITED STATES
A. ALLIGATOR RETRIEVAL DEVICEFA-88810-20

Introduction

Affected Products
  1. ALLIGATOR RETRIEVAL DEVICE
  2. X-CELERATOR EXCHANGE HYDROPHILIC GUIDEWIRE - SS
  3. ULTRAFLOW HPC FLOW DIRECTED MICRO CATHETER
  4. MARATHON FLOW DIRECTED MICRO CATHETER
  5. PIPELINE EMBOLIZATION DEVICE (25-35MM)

Reason

Medtronic has identified the potential for an issue with a specific set of lot numbers where the PTFE (polytetrafluoroethylene) coating could delaminate and separate from the delivery wire. Delamination and detachment of the PTFE coating material may lead to PTFE coating in the blood stream. PTFE in the blood stream, based on its specific size and quantity, may be safe or could lead to thromboembolic complications or irreversible injuries including but not limited to: intracranial edema, peripheral edema, incomplete treatment, infection, local inflammatory response, systemic inflammatory response, ischemic stroke, lysis/necrosis, neurological deficit, organ impairment, shock, space occupying lesion, thrombosis, and hemorrhage.


Through September 27, 2016, Medtronic has received a total of five (5) reports for the products that are potentially affected by this issue. Medtronic has received no reports of serious injuries or patient death as a result of this issue.