Health Products

NANOSTIM LEADLESS CARDIAC PACEMAKER

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.NANOSTIM LEADLESS CARDIAC PACEMAKER
Manufacturer
St. Jude Medical, Cardiac Rhythm Management Division
15900 VALLEY VIEW COURT
Sylmar
91342
California
UNITED STATES
S1DLCP

Introduction

Affected Products

Reason

This recall has been initiated due to reports of lost telemetry and pacing output as a result of a battery malfunction associated with St. Jude Medical (SJM) Nanostim Leadless Cardiac Pacemaker (LCP) devices, Model No. S1DLCP. No patient injuries have been reported in association with the loss of bradycardia pacing therapy.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime