ADVIA 560 Hematology Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
ADVIA 560 HEMATOLOGY ANALYZER11170842

Introduction

Affected Products

Reason

Siemens is investigating an issue on the ADVIA 560 Hematology System which may cause an incorrect result to be reported. Siemens has received two reports of multiple discordant records for the same sample ID (SID) in the ADVIA 560 Hematology System database that occurred during the installation of the systems. The database should only contain one record of a sample ID number for any given time and date. If there are multiple records for the same sample ID, it is possible that multiple results may be manually or automatically sent to the laboratory information system (LIS), printed or displayed on the results report screen.