C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
37305
628294
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
37305
Introduction
Affected Products
A. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
B. ACCESS SYSTEM - OSTASE ASSAY
C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
D. ACCESS OSTASE CALIBRATORS
E. ACCESS OSTASE QC
Reason
Beckman Coulter has determined that specific Access Ostase Calibrator and QC lots do not meet their expiration date claims within our 10% criteria. For these lots we determined that REF B83876 and B83877 have 68 days of stability and REF 37305 and 37309 have 51 days of stability.
Access Ostase Calibrator and QC instability could result in falsely elevated patient results or QC failures.
For 37305/37309, the instability could cause patient results to be overestimated by as much as 16% at the original expiration dates.
For B83876/B83877, the instability could cause patient results to be overestimated by as much as 16% as of October 17, 2016.