UNICEL Immunoassay System, Ostate Assay, Calibrators and QC (2016-10-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
623207
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
D. ACCESS OSTASE CALIBRATORSB83876
623208
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
E. ACCESS OSTASE QCB83877
628294
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. ACCESS SYSTEM - OSTASE ASSAY37305
628294
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY37305
628294
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY37305

Introduction

Affected Products

A. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
B. ACCESS SYSTEM - OSTASE ASSAY
C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - OSTASE ASSAY
D. ACCESS OSTASE CALIBRATORS
E. ACCESS OSTASE QC

Reason

Beckman Coulter has determined that specific Access Ostase Calibrator and QC lots do not meet their expiration date claims within our 10% criteria. For these lots we determined that REF B83876 and B83877 have 68 days of stability and REF 37305 and 37309 have 51 days of stability.

Access Ostase Calibrator and QC instability could result in falsely elevated patient results or QC failures.

For 37305/37309, the instability could cause patient results to be overestimated by as much as 16% at the original expiration dates.

For B83876/B83877, the instability could cause patient results to be overestimated by as much as 16% as of October 17, 2016.