While performing a service call, it was identified that a Somatom Definition AS was missing the laser windows on the gantry front cover.
It was identified that due to the gantry cover opening, there is a risk for severe injury of the patient if he/she is to use the opening for stabilization or for support.
Further investigation revealed that the Somatom Definition AS and the Somatom Definition Flash were affected.