Health Products

Digital Pressure Monitor

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5150402411
Manufacturer
A & D Engineering, Inc.
1756 Automation Parkway
San Jose
95131
California
UNITED STATES
Digital Pressure MonitorUA-767PBT-CI

Introduction

Affected Products

Reason

A&D may have shipped blood pressure monitors that were not labeled for use with small cuff. The device is intended to be used with arm sizes from 5.1 inches to 17.7 inches or small cuffs (6.3-9.4") , medium cuffs ( 9.4-14.2"), and large cuffs (14.2-17.7"). For optimal comfort, the manufacturer adjusts the exhaust valve for the devices to be used with a small cuff. In doing so, the manufacturer may have not provided clear labeling on the front of these devices to indicate they have been adjusted for small cuff use. Additionally, some devices may have been labeled "adjusted for small cuffs" but may have been equipped with a medium cuff.