INNOVA WASHER, DISINFECTOR AND CMS (2016-10-25) Oct 25, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer. Manufacturer BHT HYGIENETECHNIK GMBH MESSERSCHMITTSTRASSE 11 GERSTHOFEN 86368 GERMANY B. INNOVA E4 WASHER - DISINFECTOR 50600.13 59742001 Manufacturer BHT HYGIENETECHNIK GMBH MESSERSCHMITTSTRASSE 11 GERSTHOFEN 86368 GERMANY C. INNOVA E4 CMS 50600.92 4871098 Manufacturer BHT HYGIENETECHNIK GMBH MESSERSCHMITTSTRASSE 11 GERSTHOFEN 86368 GERMANY A. INNOVA M4 WASHER - DISINFECTOR 50200.01
Affected Products
INNOVA M4 WASHER - DISINFECTOR
INNOVA E4 WASHER - DISINFECTOR
INNOVA E4 CMS
The manufacturer received reports of incidents when one of the door springs broke and a piece of the broken spring struck the operator. This recall has been initiated to take a preventive action measure to improve the safety of the door springs used in the units.