This voluntary recall follows confirmation of a single user report from outside of Canada which resulted in 3M identifying an unintended material (process liner) in the product. The presence of a process liner may prevent or impede the safe return of electrosurgical current following electrosurgical unit (ESU) activation. This defect has the potential to increase the risk of a patient burn directly under the pad or at an alternate site of the body due to increased impedance at the pad application site.