NucliSENS Magnetic Silica and Extraction Reagents (2016-11-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Z017KA1MS
Manufacturer
Biomerieux Sa
Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
B. NucliSENS Magnetic Extraction Reagents280133
Z017KA1MS
Manufacturer
Biomerieux Sa
Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
A. NucliSENS easyMAG - Magnetic Silica280133

Introduction

Affected Products

A. NucliSENS easyMAG - Magnetic Silica
B. NucliSENS Magnetic Extraction Reagents

Reason

Following additional quality controls with a parameter (1 mL specimen/BK Virus) which mimic the worst case condition, a drift of the extraction performances was detected. This drift/decrease of performance after NucliSENS easyMAG, NucliSENS miniMAG, and NucliSENS eMAG Extraction induced a difference superior to -0,5 log quantification on amplification assays on DNA. This difference superior to -0,5 log is significant on results. As the downstream applications are qualitative and/or quantitative methods, if the customer does not detect the issue, the decrease of downstream application performance could lead to: 1. A risk of false negative, for qualitative tests 2. Invalid results, when the internal amplification control is not on specifications. 3. Under quantified results for quantitative tests.