Health Products

MRIDIUM MRI Infusion System - Pump

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Iradimed Corporation
1025 Willa Springs Dr.
Winter Springs
32708
Florida
UNITED STATES
MRIDIUM MRI Infusion System - Pump3860

Reason

Change to 1 (one) page on an instruction card set that attaches to the 3860 units. IRadimed Corp continues to investigate reports of possible fluid free flow, and concluded that stretching or improper placement of the IV set during pump set up is the primary cause of these reports. In 2012, IRadimed customers received an updated quick reference instruction card which emphasized proper placement of IV set during pump set up. Since that time, newer customers did not receive this updated quick reference instruction card. Though no incidences of injury have occured, IRadimed is proactively re-issuing this notice to assure the continued safe use of MRidium MR Infusion Pump Systems.