Health Products

Central Water Plant Reverse Osmosis Units (CWP 100 and 60) (2016-11-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. Central Water Plant Reverse Osmosis Unit (CWP 60)
Manufacturer
Gambro Lundia AB
MAGISTRATSVAGEN 16, P.O. BOX 10101,
Lund
22010
SWEDEN
WRO63
More than 10 numbers, contact manufacturer.A. Central Water Plant Reverse Osmosis Unit (CWP 100)
Manufacturer
Gambro Lundia AB
MAGISTRATSVAGEN 16, P.O. BOX 10101,
Lund
22010
SWEDEN
WRO 104S

Introduction

Affected Products

A. Central Water Plant Reverse Osmosis Unit (CWP 100)
B. Central Water Plant Reverse Osmosis Unit (CWP 60)

Reason

Baxter Corporation is issuing a device correction for specific models of the central water plant (CWP) reverse osmosis units in order to correct the insufficient crimping of the cables within the main pump. The issue was identified as a result of customer complaints received for the presence of smoke in the water treatment room due to the overheating of the pump in the CWP. As part of Baxter's investigation, based on pictures provided from the event, the most likely cause of the thermal event was due to the crimp of the cables for the main pump being insufficient. To minimize any potential side effects of thermal events, please ensure all panels of the CWP are in place.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime