Health Products

LigaSure Blunt Tip Laparascopic and Impact Open Sealer/Divider (2016-11-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS
1810 West Drake Dr.
Tempe
85283
UNITED STATES
B. LigaSure Impact Open Sealer/DividerLF4318
More than 10 numbers, contact manufacturer.
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS
1810 West Drake Dr.
Tempe
85283
UNITED STATES
A. LigaSure Blunt Tip Laparoscopic Sealer/DividerLF1637

Introduction

Affected Products
  1. LigaSure Blunt Tip Laparoscopic Sealer/Divider
  2. LigaSure Impact Open Sealer/Divider

Reason

Manufacturer has received complaints reporting that reprocessed LF1637 or LF4318 Ligasures experienced compatibility issues when connected to the FT10 generator. An "invalid instrument" error message may appear on FT10 generators with software revision 1.1.5.4 when devices without a radio-frequency identification (RFID) chip are connected to the generator. Updates were made to the instructions for use (IFU) to provide clarity on software compatibility.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime