Health Products

DYND70000 Medicine Cup 30 mL

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lot numbers before 122212515DYND70000 Medicine Cup 30 mL
Manufacturer
Medline Industries Inc
1 Medline Place
Mundelein
60060
Illinois
UNITED STATES
DYND70000

Reason

The medicine cups have gradation markings on the side in millilitres. The increments are 2.5, 5, 7.5, 10, 12.5, 15, 20, 25, 30 mL. Some cups repeat the 5 mL mark at the 10 mL location so that the increments appear as: 2.5, 5, 7.5, 5, 12.5, 15, 20, 25, 30 mL. Therefore a certain number of cups in each case of product will have incorrect markings. The cups are injection molded on a machine that has 64 cavities (a cavity is a mold for a cup). Therefore 64 cups can be produced at once with each injection of material. Each cavity is numbered and the number is embossed on the bottom of the cup. Cavity #44 is the defective mold with the incorrect increments. Therefore every cup numbered #44 on the bottom will have the defect. The cups are packaged in sleeves of 100 stacked cups and packaged 500 stacks in a case (5,000 cups). There will be 2, 4, or 6 sleeves containing cups from cavity #44 in a case, which comes out to 50-150 cups per case. This is based on how the robot stacks, combines, and bags the product. Therefore there is the potential for 150 defective cups in a case.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime