Abbott has identified a lot specific stability issue with both reagent lots 10533UP00 and 10534UP00. As the reagent ages, relative light unit (RLU) values for the calibration curve will decrease causing patient results calculated from this depressed curve to over-recover, potentially resulting in falsely elevated patient results. Controls may not reflect reagent instability as recalibration may result in control values within acceptable ranges.