Health Products

Semislip Guidewire

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
C15-4089Semislip Guidewire
Manufacturer
UROTECH GmbH
MEDI-GLOBE-STRASSE 1-5
ACHENMUHLE
68305
GERMANY
UAS-35150-S

Reason

Because of a lack in strength of the material, there is a potential for the hydrophilic-coated Pebax tip to loosen from the core of the guidewire which could result in some fragments of the tip of the guidewire to become loose and stay in the body of the patient. This would require surgery to remove the parts of the guidewire from the body. This problem occurred only once in a hospital in Kufstein, Austria.