Health Products

Merge Eye Station

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
10.6
Manufacturer
Merge Healthcare Incorporated,
900 Walnut Ridge Drive,
Hartland
53029
Wisconsin
UNITED STATES
Merge Eye StationMERGE EYE STATION

Reason

This recall has been initiated due to an issue related to the potential accidental deletion of record(s) by an eye station user.  In the current workflow, a user can select to delete a record(s).  When that selection is made, a confirmation box is presented asking the user if they want to delete the record(s). The user can then either left click on "ok" or press the "enter" key to confirm the action. Currently, this confirmation defaults to "ok" or "yes". In this workflow, a user could inadvertently hit the "enter" key, which would delete the images or procedures. Once these images or procedures are deleted, they are lost permanently with no recovery.