Health Products

Shockpulse Lithotropsy Generator

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
CG0032
Manufacturer
Cybersonics Inc.
5340 Fryling Road, Suite 101
Erie
16510
Pennsylvania
UNITED STATES
Shockpulse Lithotropsy GeneratorSPL-G

Reason

Revised labeling- the rating label on the back of the ultrasonic generator specifies a voltage range of 100V to 240V , and based on tests this needs to be revised to 100V to 150V.