Ceruloplasmin for Roche/Hitachi COBAS C Analyzers

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
19395501
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Ceruloplasmin for Roche/Hitachi COBAS C Analyzers20764663322

Reason

During an internal investigation the Ceruloplasmin Assay (Cat 20764663322 Lot 19395501/only lot imported in Canada) will need to have the sensitivity limit parameter changed when used on the Cobas C311/C501/C502/C701/C702 until further notice. Higher signals could be generated causing problems with the upper sensitivity limit that could generate error alarms that would prevent performing any measurements.