Health Products

Poche A Sang Quadruple

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not Applicable
Manufacturer
Macopharma
Rue Lorthiois
Mouvaux
59420
FRANCE
Poche A Sang QuadrupleLQT7292LX

Reason

Macopharma was informed of a number of fast filtrations on the NPT6280LA reference in use in France. This has caused an absence of effective filtration through the media of the LCRD2 Filters. This sporadic defect is linked to a partial absence of the weld in the media caused by poor positioning (fold or gap) of the last polyester layer in the mattress.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime