Health Products

Laerdal Compact Suction Unit 4 (LCSU 4)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
150505A0836 to 160901A0226Laerdal Compact Suction Unit 4 (LCSU 4)
Manufacturer
Laerdal Medical AS
Tanke Svilandsgate 30, P.O. Box 377,
Stavanger, Rogaland Fylke
4002
NORWAY
88005101

Reason

The on/off functionality in the LCSU 4 is controlled by electronic components. One of these components might be damaged. The damage occurs over time when the unit is connected to a power source (battery or external power source), but is not related to frequency of use.
A damaged component will result in failure of the on/off functionality: the LCSU 4 cannot be turned off, and it may turn on by itself.