Artis Devices (2016-12-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
F. Artis Q.Zen10848353
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
D. Artis Zee Biplane10094141
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
E. Artis Q10848280
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
B. Artis Zee Multi-Purpose10094139
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
C. Artis Zee Ceiling10094137
All lots
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
A. Artis Zeego Multi-Axis10280959

Introduction

Affected Products

A. Artis Zeego Multi-Axis
B. Artis Zee Multi-Purpose
C. Artis Zee Ceiling
D. Artis Zee Biplane
E. Artis Q
F. Artis Q.Zen

Reason

1. In Artis Systems with A100Plus or A100G Generators, an attempt to resume operation following detection of fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator.

2. For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with the graphics card may in rare cases result in the loss of image display in the examination room.