UreteroReno Fiberscope

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers
Manufacturer
Olympus Medical Systems Corp.
2951 Ishikawa-cho
Hachioji-shi, Tokyo
192-8507
JAPAN
UreteroReno FiberscopeURF-P6

Reason

The manufacturer, Olympus Medical Systems Corp. (OMSC), has initiated this corrective action following complaints on the URF-P6/P6R insertion tube, and breaks of the bending tube. These complaints have not resulted in any known adverse event. The URF-P6/P6R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). OMSC is aware of adverse events on the URF-V2/V2R endoscopes which have a similar structure to the URF-P6/P6R endoscopes. Olympus is providing a two-page instructions for safe use to remind customers of inspection instructions in the existing operation manual and also provide pictures and instructions to assist in performing the inspection.