C. Fresenius 2008K2 Fifth Module Dialysate Delivery System
190610
All serial numbers
Manufacturer
Fresenius USA Inc.
4040 Nelson Avenue
Concord
94520
California
UNITED STATES
A. Fresenius 2008K Dialysate Delivery System
190300
All serial numbers
Manufacturer
Fresenius USA Inc.
4040 Nelson Avenue
Concord
94520
California
UNITED STATES
J. Fresenius 2008K2 Dialysate Delivery System
R190395
Introduction
Affected Products
Fresenius 2008K Dialysate Delivery System
Fresenius 2008K2 Dialysate Delivery System
Fresenius 2008K2 Fifth Module Dialysate Delivery System
K@Home
Fresenius 2008T Hemodialysis System
Fresenius 2008K2@Home
Fresenius 2008T Hemodialysis System CDX
2008T Hemodialysis System CDX - Refurbished
2008K@Home - Refurbished
Fresenius 2008K2 Dialysate Delivery System
Reason
The purpose of this recall is to inform users of the 2008 series hemodialysis machines of a software issue pertaining to the ultrafiltration (UF) rate during treatment. It was discovered that when the UF rate, goal or time is adjusted using the "up" and "down" arrow keys of the hemodialysis machine and the change is cancelled using the "escape" key without pressing the "confirm" key first, the "cancelled" UF rate is actually being executed rather than the rate displayed on the machine.