Health Products

NX 3.0.8950 Image Processing Software

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Software Version NX 3.0.8950NX 3.0.8950 Image Processing Software
Manufacturer
AGFA Healthcare N.V.
Septestraat 27
Mortsel
2640
BELGIUM
8900

Reason

Due to a memory issue a warning message about unexpected error and possible incorrect data integrity is shown to the user. A reboot of the workstation computer after having received the warning message prevents the issue from occurring. In one instance the warning instructions were not followed and resulted in the image of one patient being archived in another patient's record.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime