Health Products

S-ICD System - Q-Tech Programmer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Cardiac Pacemakers Incorporated
4100 Hamline Avenue North
St. Paul
55112
UNITED STATES
S-ICD System - Q-Tech Programmer3200

Reason

The potential for radio frequency (RF) interference to alter wireless communication from a programmer, which in rare instances may cause an S-CID to perform an unintended command. This behavior can only occur during an active, in-clinic interrogation/programming session with the Model 3200 S-ICD programmer. There is no risk of this behavior occurring when the latitude patient management system communicates with an S-ICD in an ambulatory setting.