Health Products

Apollo

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
08060167
Manufacturer
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY
Apollo9784020231

Reason

Villa Sistemi Medicali became aware that a patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and the table was moving from vertical position.
The analysis showed that the footrest of the device detached from the table top because the safety screws were discarded at installation. The purpose of these screws is to prevent the footrest to completely detach from the table in order to avoid the patient to fall.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime