V60 Ventilator, Symbols Only, With Cflex And Avaps Options

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Respironics California, Inc.
2271 Cosmos Court
Carlsbad
92011
California
UNITED STATES
V60 Ventilator, Symbols Only, With Cflex And Avaps Options1053614

Reason

The V60 ventilator with version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent INOP) and display error code 100E. Ventilatory support will cease.

An audible high-priority alarm will sound continuously for at least 2 minutes when the V60 shuts down for any vent INOP condition and is operating on battery power. If the V60 is connected to AC power (mains supply), the alarm will continue to sound until an operator intervenes. If the V60 is connected to a remote alarm system, the alarm system will be activated until action is taken by the operator.

Error code 100E only exists in V60 version 2.20 software. Therefore, V60's running version 2.10 software are not subject to this particular vent INOP condition. An audible high-priority alarm will sound continuously for at least 2 minutes when the V60 shuts down for any vent INOP condition and is operating on battery power. If the V60 is connected to AC power (mains supply), the alarm will continue to sound until an operator intervenes. If the V60 is connected to a remote alarm system, the alarm system will be activated until action is taken by the operator.

Error code 100E only exists in V60 version 2.20 software. Therefore, V60's running version 2.10 software are not subject to this particular vent INOP condition.