Health Products

OmniLab V.12.0 - Laboratory Information System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Version 12.10.0.116OmniLab V.12.0 - Laboratory Information System
Manufacturer
Omnitech Labs Inc.
Suite 1305, 215 Boul. Du Seminaire Sud
SAINT-JEAN-SUR-RICHELIEU, Quebec
J3B 8W1
CANADA
VERSION 12.10.0.116

Reason

While investigating a problem reported by a client, the Omnitech Labs team discovered that a combination of configurations prevents microbiology cases with a status of "preliminary amendment" and "preliminary comment" from appearing on the lists of cases awaiting processing. In short, the problem affects clients whose systems are configured for the use of validation by a microbiologist and the "confirmation" button that is found on the "identification" tab.