Health Products

RX Daytona System - Lipase

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Randox Laboratories Ltd.
55 Diamond Road
Ardmore, County Antrim
BT29 4QY
UNITED KINGDOM
RX Daytona System - LipaseLI3837

Reason

Reagent probe carryover has been observed from triglycerides and direct LDL (low density lipoprotein) reagents to lipase reagent on the RX Daytona, RX Daytona+ and RX Imola Analysers causing elevated lipase test results. It occurs when the lipase is tested directly before or after triglycerides or direct LDL in the test running order. It is also possible that this may occur on other instrumentation.