Health Products

BrightView Imaging Systems (2017-02-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
B. BrightView XCT Imaging System453560462131
All lots
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
C. Sys, BrightView X with Crystal, 3/8" (Upgradeable)453560829241
All lots
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
BrightView Gamma Camera System2170-3002A

Introduction

Affected Products
  1. BrightView Gamma Camera System
  2. BrightView XCT Imaging System
  3. Sys, BrightView X with Crystal, 3/8" (Upgradeable)

Reason

The detector braking systems may be rendered ineffective resulting in one or both detectors moving to their mechanical limit. In the case of this kind of failure, the detector motion cannot be halted using normal operator intervention (e-stop, collimator contact sensor activation).