Health Products

Brainscan Radiosurgery Systems (2016-11-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
iPlan RT Dose 4.1.1
Manufacturer
BrainLab AG
Olof-Palme-Strasse 9
Munich
81829
GERMANY
D. iPlan RT Dose21213E
BRAINSCAN PLATFORM SW 5.31
Manufacturer
BrainLab AG
Olof-Palme-Strasse 9
Munich
81829
GERMANY
B. Brainscan Radiosurgery System - Software20140B (5.31)
BRAINSCAN HARDWARE
Manufacturer
BrainLab AG
Olof-Palme-Strasse 9
Munich
81829
GERMANY
C. Brainscan Radiosurgery System - Head Ring40951-1
IPLAN RT DOSE 3.0
Manufacturer
BrainLab AG
Olof-Palme-Strasse 9
Munich
81829
GERMANY
A. Brainscan Radiosurgery System - iPlan RT Dose 3.0 Platform Software21213

Introduction

Affected Products

A. Brainscan Radiosurgery System - iPlan RT Dose 3.0 Platform Software
B. Brainscan Radiosurgery System - Software
C. Brainscan Radiosurgery System - Head Ring
D. iPlan RT Dose

Reason

Brainlab Radiation Treatment Planning Software: Potential inaccurate dose calculation for small multileaf collimator(MLC)-shaped fields when using the pencil beam algorithm.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime