Health Products

Cobas IT Middleware

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All versions
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Cobas IT Middleware06498027001

Reason

In specific circumstances, the evaluation of QC Westgard rules can be performed incorrectly. Rules comparing a quality control result with a previous result (or results) can be incorrectly assessed out of sequence. Rules to assess a rising or falling trend in quality control results can also be wrongly evaluated. However, the result released is correct.