Lot or serial numberCompaniesBrand nameModel or catalog number
15040308
Manufacturer
Omnia Spa
Via Francesco Delnevo 190sx
Fidenza PR
43036
ITALY
Irrigation Set32.F0134.00

Reason

Omnia has determined that during the assembling phase, if tightening rings are not perfectly placed, it is possible that in a condition of maximum flow and pressure the tube can detach causing leakage of saline solution.