Heartstart Mrx Defibrillator/Monitor (2017-02-27)
· Published by Health Canada · Official notice
Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
| Lot or serial number | Brand name | Companies | Model or catalog number |
|---|
| All lots | B. Heartstart Mrx Defibrillator/Monitor | - Manufacturer
- Philips Medical Systems
- 3000 Minuteman Road
- Andover
- 01810
- Massachusetts
- UNITED STATES
| M3535A |
| All lots | A. Heartstart Mrx Defibrillator/Monitor | - Manufacturer
- Philips Medical Systems
- 3000 Minuteman Road
- Andover
- 01810
- Massachusetts
- UNITED STATES
| M3536A |
Affected Products
A. Heartstart Mrx Defibrillator/Monitor
B. Heartstart Mrx Defibrillator/Monitor
Philips has become aware of the following issues:
1. If the pins that connect the battery to the device have become damaged or full of debris, the device may not power on or may repetitively reboot.
2. When the device is unplugged from AC mains, it may enter a state where it will not charge for shock delivery or may unexpectedly cease pacing.