Health Products

Heartstart Mrx Defibrillator/Monitor (2017-02-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsB. Heartstart Mrx Defibrillator/Monitor
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
M3535A
All lotsA. Heartstart Mrx Defibrillator/Monitor
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
M3536A

Introduction

Affected Products

A. Heartstart Mrx Defibrillator/Monitor
B. Heartstart Mrx Defibrillator/Monitor

Reason

Philips has become aware of the following issues:

1. If the pins that connect the battery to the device have become damaged or full of debris, the device may not power on or may repetitively reboot.

2. When the device is unplugged from AC mains, it may enter a state where it will not charge for shock delivery or may unexpectedly cease pacing.