Health Products

IsoAir Mattress

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
15C0036IsoAir Mattress
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
UNITED STATES
2940-000-038

Reason

Stryker is the distributor of the IsoAir system. The specification developer/manufacturer of the IsoAir system has determined that electrical interference emanating from the system may cause clinically unacceptable levels of signal artifact in electrophysiological monitoring devices, e.g. ECG, EKG or EEG. This only occurs when the active sensor technology (AST) cable from the air pump is connected to the surface. Stryker has received a few complaints associated with this issue.