Medtronic Model 8870 N'Vision Application Card (2017-03-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
F. N'Vision Application Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
G. RestoreSensor Multi-Program Rechargeable NeuroStimulator-N'Vision Appl Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
D. Activa PC INS - N'Vision Application Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
E. Activa RC NeuroStimulator - N'Vision Application Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
B. RestoreAdvanced Multi-Program Neurostimulator - N'Vision Application Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
C. RestoreUltra Multi-Program NeuroStimulator Syst - N'Vision Application Card8870
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
A. PrimeAdvanced Multi-Program NeuroStimulator - N'Vision Application Card8870

Introduction

Affected Products
  1. PrimeAdvanced Multi-Program NeuroStimulator - N'Vision Application Card
  2. RestoreAdvanced Multi-Program Neurostimulator - N'Vision Application Card
  3. RestoreUltra Multi-Program NeuroStimulator Syst - N'Vision Application Card
  4. Activa PC INS - N'Vision Application Card
  5. Activa RC NeuroStimulator - N'Vision Application Card
  6. N'Vision Application Card
  7. RestoreSensor Multi-Program Rechargeable NeuroStimulator-N'Vision Appl Card

Reason

Medtronic is updating the Model 8870 software application card (to version BBUO1) as well as the SynchroMed Infusion System labeling to address the unintended delivery of drug during the priming bolus function. The SynchroMed priming bolus function is intended to quickly advance drug from the pump reservoir to the catheter tip to allow for therapy initiation while the patient remains under medical supervision.

The updated 8870 software card mitigates the potential for clinically relevant effects of over-delivery of unintended drug, such as respiratory depression, loss of consciousness, or death, during the full system priming bolus procedure. The therapy applications on the software card for deep brain stimulation and spinal cord stimulation remain unchanged.

This recall affects all 8870 software application cards, which are used in the 8840 N'Vision Clinical Programmers. The 8870 software application card is listed under multiple licence numbers which pertain to specific sets of devices that work together.