A. Ophthalmic Capnoxygen Mask
B. OxyArm-OxyMask Model - RTM
C. OxyArm-OxyPlus
Reason
On February 14, Southmedic received a customer complaint indicating a potential occlusion of the CO2 sampling line on a Southmedic Capnography Mask. Inspection of current stock of the lot number in question did not identify any issues however investigation into additional lots manufactured with the same subassembly resulted in identification of occluded CO2 sampling lines. Subsequently all stock of various product stock keeping units (SKUs) manufactured with a CO2 sampling line were placed on hold for further inspection.