Dimension Systems-Ammonia Assays (2017-03-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension Xpand System-Ammonia (Amm) Assay10711991
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
C. Dimension Exl System-Ammonia (Amm) Assay10711991
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Exl System-Ammonia (Amm) Assay10711991

Introduction

Affected Products

A. Dimension Exl System-Ammonia (Amm) Assay
B. Dimension Xpand System-Ammonia (Amm) Assay
C. Dimension Exl System-Ammonia (Amm) Assay

Reason

In September 2014, Siemens Healthcare Diagnostics issued customer notification 14-62 that addressed a high rate of "abnormal assay" test report messages for its Dimension Ammonia (AMM) Assay. With this letter, Siemens is reducing the AMM open well stability from 3 days to 1 day (24 hours) and providing instructions for managing this change. Dimension AMM instructions for use (IFU) revision E will reflect this change.
The new version of dimension software, 10.2, will do the following: 
  • automatically move to a new well every day. This decrease in the open well maximum duration reduces the amount of time the AMM reagent is exposed to any potential contaminates, reduces the need for customers to manually move to a new reagent well, and aids in the prevention of erroneous flagging.
  • contain the below mean factor limit of 0.90.
Siemens is working to make Dimension Software10.2 available to all customers as quickly as possible.