Health Products

EDS Light Suspension, Central Axis For Visum Led Surgical Light

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0682-001-229
Manufacturer
Stryker Communications
1410 Lakeside Pkwy #100
Flower Mound
75028
Texas
UNITED STATES
EDS Light Suspension, Central Axis For Visum Led Surgical Light0682-001-229

Reason

Stryker has become aware that there is a remote likelihood that the EDS Light Suspension, Central Axis may have an insufficiently assembled joint in the suspension. The EDS Systems have redundant connections throughout the suspensions. If the EDS Light Suspension, Central Axis falls, it could potentially cause serious injuries to healthcare providers and/or patients. To date, there have been no adverse events reported for this issue.