Health Products

EliA Sample Diluent

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
JK5NEliA Sample Diluent
Manufacturer
PHADIA AB
RAPSGATAN 7
UPPSALA
SWEDEN
83-1023-01

Reason

There have been reports of instrument malfunctions for Phadia 250 due to deformed bottles in the below mentioned products and lots.  The deformation of the bottle can cause an erroneous volume detection by the instrument which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples.  The deformation is described as a flange or brim, located on the inside lower part of the vial opening.  The instrument malfunction due to deformed bottles could cause erroneous test results, mainly false positive, with regard to all Elia assays, and mainly falsely increased test results for ImmunoCAP Total IgE and ImmunoCAP IgG4/IgG assays.  The frequency of this bottle defect is very low and not every defective bottle will cause the error.  An instrument malfunction caused by deformed bottles will generate the instrument error message "3-145 right arm liquid detection below lower limit with target 2".  It is important that you review your records for the above mentioned error message.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime