ITS4850A Intellivue Telemetry System-Main Unit

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
ITS4850A Intellivue Telemetry System-Main UnitITS4850A

Reason

Philips has recently discovered that under specific 802.11 network conditions, a partial re-association of the MX40 WLAN monitor to a compatible central monitoring system ("information center") is possible. In this state, although the information center provides a visible and audible "no data tele" INOP alert, the MX40 WLAN itself enters telemetry mode, i.e., its screen turns off in one minute and local alarming is disabled.