Engen Laboratory Automation System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
J16XQH4JEngen Laboratory Automation System
Manufacturer
Thermo Fisher Scientific Oy
P.O. Box 100, Ratastie 2
Vantaa
01621
FINLAND
ENGEN

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall in regard to a software anomaly where the TCAutomation Software may send incorrect information to a VITROS system through the InOut communication interface.  This event can cause an unintended sample to be moved into the bypass aspiration position and aspirated, and results to be mis-associated with the wrong patient sample ID without the operator being notified. The InOut communication interface is currently used by all Engen Systems.