Dimension Vista Intelligent Lab Systems (2017-03-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10, contact manufacturerB. Dimension Vista 500 Intelligent Lab System
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
1000034648
More than 10, contact manufacturerA. Dimension Vista 1500 Intelligent Lab System
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
1000033793

Introduction

Affected Products

A. Dimension Vista 1500 Intelligent Lab System
B. Dimension Vista 500 Intelligent Lab System

Reason

In isolated cases when ECREA is processed immediately after the weekly automated acid clean routine during probe test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result. Investigation has determined that the processing of ECREA quality control after the probe test will identify a potential elevated ECREA result.