Advia Centaur Systems - Br (27.29) Asssays (2017-03-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsB. Advia Centaur Cp System - Br (27.29) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
3896216
All lotsA. Advia Centaur System - Br (27.29) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
3896216

Introduction

Affected Products

A. Advia Centaur System - Br (27.29) Assay
B. Advia Centaur Cp System - Br (27.29) Assay

Reason

Siemens Healthcare Diagnostics has confirmed that the Advia Centaur BR (ca 27.29) assay does not dilute linearly when used with multi-diluent 1 kit lots ending in 2577 and all future lots until further notice. When diluted, some samples over-recover the expected neat value by greater than 120%.