Advia Centaur / XP / XPT

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
20739217
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
Advia Centaur / XP / XPT02419937

Reason

Potential for the Advia Centaur Syphilis assay to interfere with the Br (Ca 27.29) assay on the Advia Centaur Systems. This interference only occurs when the Syphilis assay immediately precedes the Br assay on the Advia Centaur, Advia Centaur XP or Advia Centaur XPT systems. The Advia Centaur CP system is not impacted. The issue affects all current and future lots of Advia Centaur Systems Br reagents until a solution is implemented.