Health Products

CADD Administration Set

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
46x1001
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES
CADD Administration Set21-7339-24

Reason

A drawing on the IFU incorrectly represents the actual filter size of the contents. Where the drawing should indicate that the package contains an administration set with a 0.2 micron filter, it is incorrectly illustrated as a 2.0 micron filter. The actual product in the package and the text based description on the IFU correctly conformed to the 0.2 micron filter specification for P/N 21-7339-24. It was the filter label on the drawing that was incorrect.