Allura XPER Systems (2017-03-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
F. Allura XPER FD10 (Product of Alluraclarity Family)722026
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
G. Allura XPER FD20/20 (Product of Alluraclarity Family)722038
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
D. Allura XPER FD20 System-Main UnitALLURA XPER FD20
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
E. Allura XPER FD20 OR Table1459
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
B. Allura XPER FD20/10 System-Main UnitALLURA XPER FD20/10
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
C. Allura XPER FD10ALLURA XPER FD10
All lots
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
A. Allura XPER FD10/10-Main UnitALLURA XPER FD10/10

Introduction

Affected Products
  1.  Allura XPER FD10/10-Main Unit
  2.  Allura XPER FD20/10 System-Main Unit
  3.  Allura XPER FD10
  4.  Allura XPER FD20 System-Main Unit
  5.  Allura XPER FD20 OR Table
  6. Allura XPER FD10 (Product of Alluraclarity Family)
  7. Allura XPER FD20/20 (Product of Alluraclarity Family)

Reason

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the r-cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.